DENTAL EQUIPMENT --->

How to Maintain your Equipment

     
 

Handpieces

 
     
     
 

Important Information on Handpiece Maintenance

Is your handpiece lubricant causing your handpieces to freeze up prematurely?
Some lubricants break down in autoclave leaving a residue build up on the bearings causing them to freeze up or run sluggish.
Try using a High Temperature Autoclavable Handpiece lubricant like "STER-L".

 

How do I Clean & Maintain Handpieces

 

Between Patient Handpiece Maintenance
  Lubricate Handpiece

Place 2-3 drops of oil into the intake tube located at the bottom of the handpiece. Then operate the handpiece, with a bur installed, for approximately 10 seconds to expell the oil through and out of the handpiece. Doing this will flush the debris and contaminations from the handpiece. Warning: Operating the handpiece without a bur installed will result in premature chuck failure.

  Clean Exterior of Handpiece

Wipe down the exterior of handpiece with alcohol or distilled water. Remove gross contaminates with a fine wire brush. Do not use tap water or phosphate based soaps.

  Sterilize

Individually bag each handpiece for sterilization. Do not exceed 135C/275F. If using a chemical sterilizer, use clean chemicals. After cycle allow handpiece  to return to room temperature before use, do not operate that is hot to touch. Warning: Never leave a handpiece in the sterilizer overnight.

  Weekly Maintenance

Clean the chuck of the handpiece once a week. Spray handpiece cleaner up into the front of the chuck. Removing debris from the chuck assembly will prolong the life of the chuck.

Between Patient Handpiece Maintenance [Lube Free]
  Clean Exterior of Handpiece

Wipe down the exterior of handpiece with alcohol or distilled water. Remove gross contaminates with a fine wire brush. Do not use tap water or phosphate based soaps.

  Sterilize

Individually bag each handpiece for sterilization. Do not exceed 135C/275F. If using a chemical sterilizer, use clean chemicals. After cycle allow handpiece to return to room temperature before use, do not operate a handpiece that is hot to touch. Warning: Never leave a handpiece in the sterilizer overnight.

  Weekly Maintenance

Clean the Chuck of the handpiece once a week. Spray handpiece cleaner up into the front of the chuck. Removing debris from the chuck assembly will prolong the life of the chuck.

Note:

There are many ideas expressed in the dental industry in regards to which form of handpiece maintenance will yield the longest handpiece life. We are not representing that following this maintenance procedure outline will guarantee an increase in handpiece life. The maintenance procedure expressed above is what we have found to offer the best results in the tests we have conducted.

Source: HPP Inc.

 

 
 

 

 

You are visitor no.
Dental Equipment Site Counter
 